Download Flyer
PILATUS 2M detector
PILATUS 2M prototype at the X12SA coherent small-angle X-ray scattering beamline, SLS
Ultimate speed and data quality for small-angle X-ray scattering

The PILATUS 2M detector is perfectly suited for modern small-angle X-ray scattering (SAXS) and macromolecular crystallography (MX) beamlines, which require fast detectors with very high dynamic range.

It is based on the newly developed CMOS hybrid-pixel technology and operates in single-photon-counting mode. With its large active area, high dynamic range, high frame rate, the PILATUS 2M is ideally suited for SAXS.

The detector system operates at room temperature, with a cooling unit for temperature stabilization. It comes as a complete system with detector, water cooler, a high-end server to cope with the high data transfer rates and the data acquisition and analysis software package TVX.

The PILATUS 2M is delivered fully calibrated and ready to use.


Applications:

  • X-ray diffraction (XRD)
  • Small-angle scattering (SAXS)
  • Macromolecular crystallography (MX)


Technical specifications

Number of modules 3 x 8 = 24
Sensor Reverse-biased silicon diode array
Sensor thickness 320 µm
Pixel size 172 x 172 µm2
Format 1475 x 1679 = 2,476,525 pixels
Area 254 x 289 mm2
Intermodule gap x: 7 pixels, y: 17 pixels, 8.0% of total area
Dynamic range 20 bits (1:1,048,576)
Counting rate per pixel > 2 x 106 X-ray/s
Energy range 3 – 30 keV
Quantum efficiency
(calculated)
3 keV: 80%
8 keV: 99%
15 keV: 55%
Energy resolution 500 eV
Adjustable threshold range 2 – 20 keV
Threshold dispersion 50 eV
Readout time 3.6 ms
Framing rate 30 Hz
Point-spread function 1 pixel
Data formats Raw data, TIF, EDF, CBF
External trigger/gate 5V TTL, 3 different modes
Software interface Through socket connection; clients for EPICS, SPEC and stand-alone operation are available
Cooling Close-circuit water cooling unit for temperature stabilization
Power consumption 200 W
Dimensions (WHD) 384 x 424 x 458 mm
Weight 42 kg